Introduction: Measuring blood pressure is a habitual practice in community pharmacies as well as in patients’ homes. Therefore, it is important to have a technique to verify the reliability and validity of blood pressure monitors, both the ones in the pharmacies as well as the ones that users have at home.
Material and methods: For this, we propose and carry out a practical case of assessing reliability using the Bland and Altman technique, and of assessing validity by means of a t-test for matched data and an adaptation of the criteria of the European Society of Hypertension (ESH).
Results: In the practical case we deal with, the monitor assessed was found to be reliable in comparison with our reference monitor with regard to the SBP as well as the DBP, showing a bias in the DBP, measuring on average 5.1 mmHg more than the reference monitor. The monitor assessed did not pass, with regard to ours, the ESH’s validation criteria.
Discussion: As a personal contribution, an Excel® spreadsheet was created that did all the above-mentioned calculations, including the t-test power, as well as the Bland and Altman graphs on introducing the results of the monitoring in question.